5. Click “Calibrate” in order to calibrate the whole scanner.
In this step the software will first measures the position, orientation, focal length and distortion
coefficients of the camera. Then it will project a pattern sequence in order to determine the same
optical properties of the projector. If Texturing is not switched off, it will also project different colors
to perform a white balance measurement.
After a successful calibration, DAVID will project a checkerboard pattern, whose corner points should fall
exactly onto the calibration markers.
Your scanner is now calibrated. This refers to the position and orientation of camera and projector relative to
each other, and the focus and brightness settings. You may move, rotate and tilt the scanner as a whole, and
you can close and restart the DAVID software, without losing the calibration. You may also change the
Projector Brightness setting in order to adjust image brightness (red sine curves) to the scanned object.
However, if you rotate camera or projector with respect to each other, or change the focus (e.g. in order to
scan much smaller or larger objects), you will have to repeat the whole calibration procedure.